SPIE Medical Imaging 2012: Image-Guided Procedures, Robotic Interventions, and Modeling, Vol. 8316, pp. 83162V, 2012

Real-Time Circumferential Mapping Catheter Tracking for Motion Compensation in Atrial Fibrillation Ablation Procedures

Opens internal link in current windowAlexander Brost, Felix Bourier, Opens external link in new windowAndreas WimmerOpens external link in new windowMartin Koch, Atilla Kiraly, Rui Liao, Klaus Kurzidim, Opens external link in new windowJoachim Hornegger, and Norbert Strobel

Motivation

Atrial fibrillation (AFib) is the most common heart arrhythmia, and leads to an increased risk of stroke for the patients. Since the first treatment approaches using radio-frequency ablation, this method has now become an accepted treatment option, in particular when drug therapy fails. Catheter ablation procedures are performed in electrophysiology (EP) labs usually equipped with modern C-arm X-ray systems. Augmented fluoroscopy, overlaying 2-D renderings obtained from either CT, MR, or C-arm CT 3-D data sets onto live fluoroscopic images, can facilitate more precise real-time catheter navigation and also reduce radiation. Unfortunately, catheter navigation under augmented fluoroscopy is compromised by cardiac and respiratory motion. A first approach to tackle this problem by providing a motion compensated overlay was proposed Opens external link in new windowhere. It involved tracking of commonly used circumferential mapping catheters. As AFib therapy takes place in the vicinity of the circumferential mapping catheter, tracking the catheter can be assumed to capture the motion of the relevant treatment region correctly if the device has been firmly positioned. Fortunately, we can count on the physicians to provide a stable wall contact, as it is in their best interest. Otherwise complete isolation of the pulmonary veins (PVs) may fail due to undetected residual PV-atrial electrical connections. The initially proposed method generated a 3-D model of the catheter and applied an unconstrained 2-D/3-D registration to register the catheter model to biplane fluoroscopy images. This work was extended regarding robustness using a learning-based catheter segmentation. The approach was also modified to work on monoplane systems as well, see Opens external link in new windowhere. This approach does not require simultaneously biplane fluoroscopy but instead uses a constrained 2-D/3-D registration to achieve motion compensation if only monoplane acquisitions are used at a biplane system. We extended this method to achieve real-time capability, even outperforming other motion compensation approaches.

Methods

The proposed method comprises three main steps. First, a 3-D model of the mapping catheter is generated. Second, the catheter is segmented in 2-D X-ray images using a classifier. The segmentation result is refined next. The main difference to previous approaches is that our method does not use the a thinning algorithm. Finally, the catheter is tracked relying on 2-D/3-D registration of the 3-D catheter model to the 2-D X-ray images. The resulting motion information can be used to adjust the fluoroscopic overlay accordingly. Below, these steps are explained in more detail Opens external link in new windowhere.

Evaluation and Results

For the evaluation of the proposed method, 13 clinical biplane sequences were available. Biplane sequences were required to calculate an estimate for the 3-D error of our motion compensation approach. The 2-D tracking error was calculated as the average distance between the motion-compensated catheter model and the manually segmented mapping catheter. Manual segmentation was supervised by a cardiologist. This distance was averaged over all frames of a particular sequence to arrive at an overall tracking error for each sequence. Our proposed method achieved a 2-D tracking error of 0.61 mm +/- 0.45 mm. The 3-D tracking error was evaluated by comparing the 3-D motion vector obtained from our constrained motion compensation approach to a 3-D motion vector obtained from the biplane sequences. The 3-D error turned out to be 2.10 mm +/- 1.26 mm. The results are given below.

Tracking results of our new method in comparison to the reference method. (a) Twodimensional tracking error for all 26 sequences. For each sequence, the average, minimum, and maximum for the tracking error as well as the observed motion is given. On average, we observed for our new method an overall mean error of 0.61 mm, an overall minimum error of 0.08 mm, and an overall maximum error of 2.26 mm. The reference method yielded an overall mean of (b) Three-dimensional tracking error for all 26 sequences. We found an overall mean error of 2.16 mm, an overall minimum error of 0.01 mm, and an overall maximum error of 12.71 mm.
Discussion and Conclusions

We developed a method for respiratory and cardiac motion compensation for use in radiofrequency catheter ablation of atrial fibrillation that is capable of real-time processing. It outperforms a different approach for motion compensation and its accuracy. Although the accuracy is slightly lower than other methods, see Opens external link in new windowhere, the speed-up might be the key for its application as this method does not put any restrictions on the procedure regarding the acquisition frame rate. Excluding the skeletonization from the image processing pipeline has two effects. First, the processing speed increases, but at the same time, the method becomes less accurate. The difference between our method and the reference method is at most 0.74 mm or ∼ 4 pixels in the fluoroscopic image. The presented method appears to have a rather high 3-D error, however this is because we try to estimate 3-D motion from a single X-ray view. It is possible to reduce the 3-D error by resorting to simultaneous biplane X-ray imaging, see Opens external link in new windowhere, but this increases the X-ray dose for patient and physician. We consider our method as another step towards a fully integrated software guidance application for  electrophysiology procedures, especially atrial fibrillation ablation procedures, that accommodates physicians' needs as much as possible. Our implementation of the reference method, see Opens external link in new windowhere, was found to be able to process at least 10 frames-per-second (fps). Our new approach achieved a minimum frame rate of 20 fps.

Full Paper

Full paper available Opens external link in new windowhere.